7
Performance Tests
U1604B
Frequency: 10 kHz
Period: 100 ms
Frequency: 1 MHz
Period: 1 ms
Frequency: 40 MHz
Period: 25 ns
Trigger Sensitivity
Calibrator Setting
100 mVp-p at 5 MHz
100 mVp-p at 20 MHz
100 mVp-p at 40 MHz
Multimeter Performance Verification Tests
DC Voltage Verification Test
Calibrator Voltage
600 mV
6 V
30 V
60 V
600 V
AC Voltage Verification Test
Calibrator Voltage
6 V at 1 kHz
6 V at 30 kHz
60 V at 1 kHz
60 V at 30 kHz
300 V at 50 Hz
540 V at 1 kHz
590 V at 10 kHz
Agilent U1602B/U1604B User's and Service Guide
9.63 kHz to 10.4 kHz
96.20 ms to 103.8 ms
962 kHz to 1.04 MHz
0.96 ms to 1.04 ms
38.24 MHz to 41.93 MHz
23.85 ns to 26.15 ns
Test Limits
0.8 division at 5 MHz
1 division at 20 MHz (U1602B)
1 division at 40 MHz (U1604B)
Range
Test Limit
600 mV
597.7 mV to 602.3 mV
6 V
5.977 V to 6.023 V
60 V
29.86 V to 30.14 V
60 V
59.77 V to 60.23 V
600 V
597.7 V to 602.3 V
Range
Test Limit
6 V
5.928 V to 6.072 V
6 V
5.808 V to 6.192 V
60 V
59.28 V to 60.72 V
60 V
58.08 V to 61.92 V
600 V
295.80 V to 304.20 V
600 V
533.40 V to 546.60 V
600 V
571.1 V to 608.9 V
Channel 1
Channel 2
[PASS/FAIL]
[PASS/FAIL]
[PASS/FAIL]
[PASS/FAIL]
[PASS/FAIL]
[PASS/FAIL]
Results
Results