Automatic Measurement in Data Logger Mode

The automatic measurement for minimum, maximum and average values
are available for data logging function. You can select any of these
measurements for data recording purposes.
The real- time minimum and maximum measurements are displayed along
with the recorded time. The average value is the arithmetic average value
of the last 8 readings.
To enable the function, press F3 in the respective multimeter function
submenu.

CSV Saving Function and Interval Setting

This function allows the users to directly save their data into the USB
flash drive. To enable the function, do the following:
1 Press
2 Press F4 to proceed to page 2/2 of the data logger mode.
3 To enable the CSV saving funtion, press F2 to toggle between start and
4 To change the interval settings, press F3 repetitively to scroll down the
Figure 5-3 Enabling the CSV saving function
Agilent U1602B/U1604B User's and Service Guide
to access the data logger mode.
stop options (refer to Figure 5- 3 below). You can save up to 500
kilobytes of data.
list and make the selection (refer to Figure 5- 4 below).