Diode Verification Test

1 Put the calibrator in Standby (STBY) mode.
2 Select Meter mode of the instrument, then press F2 softkey for
3 Press F1 to select diode test.
4 Connect the calibrator to the instrument's input and ground terminal as
5 Select each function and range in the order shown in Table 7- 9. Apply
6 Compare the measurement results to the corresponding test limits
Table 7-9 Diode Verification Test
148
Ohmmeter.
shown in Figure 7- 2 using a banana plug.
the values from the calibrator shown in Table 7- 9 below.
shown in table 7- 9.
Calibrator Voltage
Function
0.9 V
Diode
Min Limit
0.8812 V
Agilent U1602B/U1604B User's and Service Guide
Performance Tests
Max Limit
0.9188 V
7