Advertisement

Resistance Verification Test (2-Wire Ohms)
Calibrator Settings
600 W
6 kW
60 kW
600 kW
3 MW
6 MW
60 MW
Capacitance Verification Test
Calibrator Settings
60 nF
600 nF
6 mF
36 mF
60 mF
300 mF
Diode Verification Test
Calibrator Voltage
0.9 V
Continuity Verification Test
Calibrator Settings
30 W
75 W
156
Range
Test Limit
600 W
595.8 W to 604.2 W
6 kW
5.958 kW to 6.042 kW
60 kW
59.58 kW to 60.42 kW
600 kW
595.8 kW to 604.2 kW
6 MW
2.973 MW to 3.027 MW
6 MW
5.958 MW to 6.042 MW
60 MW
59.28 MW to 60.72 MW
Range
Test Limit
60 nF
58.68 nF to 61.32 nF
600 nF
586.8 nF to 613.2 nF
6 mF
5.868 mF to 6.132 mF
60 mF
35.16 mF to 36.84 mF
60 mF
58.68 mF to 61.32 mF
300 mF
293.4 mF to 306.6 mF
Test Limit
0.8812 V to 0.9188 V
Condition
Buzzer sounds on
Buzzer sounds off
Results
Results
[PASS/FAIL]
[PASS/FAIL]
Agilent U1602B/U1604B User's and Service Guide
Performance Tests
Results
Results
7