Performance Tests

Oscilloscope Performance Verification Tests

The performance tests verify the channel's measurement accuracy and
acquisition system to provide confidence in functionality and accuracy of
the instrument. The performance verification tests are based on the
instrument's specification as listed in "Chapter 9 Specification and
Characteristic".

Voltage Measurement Accuracy Verification Test

This test verifies the voltage measurement accuracy. In this test, you will
measure the output of a power supply using dual cursors
instrument and compare the results with reading from the multimeter.
Test limits: 50 mV/div to 100 V/div
Refer to Table 7- 1 for the equipment required for this voltage
measurement accuracy verification test. Perform the following procedures
for channel 1, then repeat the same procedure for channel 2.
1 Set up the instrument.
* Dual Y-Cursor Measurement instructions can be found on
† Full scale on all ranges is defined as 8 divisions multiplied by the V/div setting.
Agilent U1602B/U1604B User's and Service Guide
5 mV/div to 20 mV/div
• Set the Volts/div to the value stated in the first line in Table 7- 3.
3% of full scale
±
5% of full scale
±
page
71.
*
of the
135