Agilent Technologies 11645A Series User's And Service Manual
Agilent Technologies 11645A Series User's And Service Manual

Agilent Technologies 11645A Series User's And Service Manual

Waveguide verification kit

Advertisement

Quick Links

User's and Service Guide
Agilent Technologies 11645A Series
R, Q, U, V, and W
Waveguide Verification Kit
This manual applies to the 11645A series calibration devices with serial
number prefix 3014A.
Agilent Part Number: 11645-90013
Printed in USA
Print Date: November 2004
Supersedes: January 2002
© Copyright Agilent Technologies, Inc. 1995, 2002, 2004

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 11645A Series

  • Page 1 User’s and Service Guide Agilent Technologies 11645A Series R, Q, U, V, and W Waveguide Verification Kit This manual applies to the 11645A series calibration devices with serial number prefix 3014A. Agilent Part Number: 11645-90013 Printed in USA Print Date: November 2004 Supersedes: January 2002 ©...
  • Page 2 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987).
  • Page 3 Printing Copies of Documentation from the Web To print copies of documentation from the Web, download the PDF file from the Agilent web site: • Go to http://www.agilent.com. • Enter the document’s part number (located on the title page) in the Quick Search box. •...
  • Page 5: Table Of Contents

    Returning a Kit or Device to Agilent Technologies ....... . .5-4...
  • Page 6 Contents 11645A...
  • Page 7: General Information

    General Information 11645A...
  • Page 8: Verification Kit Overview

    The following verification kits provide a set of standards with known characteristics, traceable to a reference (golden) standard in Agilent Technologies calibration lab. The sets of standards are used to verify your measurement calibration and also to verify that your network analyzer system is operating within its specifications.
  • Page 9: Equipment Required But Not Supplied

    “Contacting Agilent” on page 5-5. Agilent will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company. Refer to “Returning a Kit or Device to Agilent Technologies” on page 5-4 for instructions. 11645A...
  • Page 10: Recording The Device Serial Numbers

    General Information Recording the Device Serial Numbers Recording the Device Serial Numbers In addition to the kit serial number, the devices in this kit are individually serialized (serial numbers are labeled into the body of each device). Record these serial numbers in the appropriate table.
  • Page 11 Specifications 11645A...
  • Page 12: Specifications Environmental Requirements

    Specifications Environmental Requirements Environmental Requirements Table 2-1 Environmental Requirements Parameter Required Values/Ranges Temperature +20 °C to +26 °C (+68 °F to +79 °F) Operating ±1 °C (1.8 °F) of measurement calibration temperature Error-corrected range Altitude Operating < 4,500 meters (≈15,000 feet) Storage <...
  • Page 13: Electrical Specifications

    Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities...
  • Page 14 Specifications Electrical Specifications 11645A...
  • Page 15: Use, Maintenance, And Care Of The Devices

    Use, Maintenance, and Care of the Devices 11645A...
  • Page 16: Electrostatic Discharge

    Use, Maintenance, and Care of the Devices Electrostatic Discharge Electrostatic Discharge Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets). Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged.
  • Page 17: Visual Inspection

    Use, Maintenance, and Care of the Devices Visual Inspection Visual Inspection Examine the mating surfaces first for obvious defects or damage. If necessary, cleaning should be done every time a connection is made. Metal particles or dirt may fall onto the mating surface during the connection and disconnection process.
  • Page 18: Cleaning The Mating Plane Surfaces

    Use, Maintenance, and Care of the Devices Cleaning the Mating Plane Surfaces Cleaning the Mating Plane Surfaces 1. Use Compressed Air or Nitrogen Clean mating surfaces are essential for ensuring the integrity of the waveguide. WARNING Always use protective eyewear when using compressed air or nitrogen.
  • Page 19: Connections

    Use, Maintenance, and Care of the Devices Connections Connections Good connections require a skilled operator. Instrument sensitivity and mechanical tolerances are such that slight errors in operator technique can have a significant effect on measurements and measurement uncertainties. The most common cause of measurement error is bad connections.
  • Page 20: How To Connect: Q-Band

    Use, Maintenance, and Care of the Devices Connections b. Check the flange matings for any gap. A good connection has no gaps between the connected waveguide flanges, and the waveguide walls are flush (there is no step or offset). c. Inspect the washers for symmetrical compression. This ensures that all four screws are equally torqued.
  • Page 21: Handling And Storage

    Use, Maintenance, and Care of the Devices Handling and Storage Handling and Storage • Do install the protective end caps over the mating surfaces and store the devices in the foam-lined storage case when not in use. • Do keep devices clean. •...
  • Page 22 Use, Maintenance, and Care of the Devices Handling and Storage 11645A...
  • Page 23: Performance Verification

    Performance Verification 11645A...
  • Page 24: Introduction

    These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through Agilent Technologies. 11645A...
  • Page 25: Recertification

    A list of NIST traceable numbers may be purchased upon request to be included in the calibration report. Agilent Technologies offers a Standard calibration for the recertification of the kit. For more information, contact Agilent Technologies. Refer to “Contacting Agilent” on page 5-5.
  • Page 26 Performance Verification Recertification 11645A...
  • Page 27: Troubleshooting

    Troubleshooting 11645A...
  • Page 28: Troubleshooting Process

    Troubleshooting Troubleshooting Process Troubleshooting Process This manual contains limited information about network analyzer system operation. For complete information, refer to the instrument documentation. If your network analyzer does not pass performance verification, follow the steps in Figure 5-1 to determine the cause of the failure and the correct action to take to correct the failure.
  • Page 29 Troubleshooting Troubleshooting Process Figure 5-1 Troubleshooting Flowchart 11645A...
  • Page 30: Returning A Kit Or Device To Agilent Technologies

    Returning a Kit or Device to Agilent Technologies Returning a Kit or Device to Agilent Technologies If your kit or device requires service, contact Agilent Technologies for information on where to send it. See “Contacting Agilent” on page 5 for contact information. Include a...
  • Page 31: Contacting Agilent

    Troubleshooting Contacting Agilent Contacting Agilent This information supersedes all prior HP contact information. Online assistance: www.agilent.com/find/assist Americas Brazil Canada Mexico United States (tel) 1 800 254 2440 (tel) (+55) 11 3351 7012 (tel) 888 447 7378 (tel) 800 829 4444 (fax) 1 800 254 4222 (fax) (+55) 11 3351 7024 (fax) 905 282 6495...
  • Page 32 Troubleshooting Contacting Agilent 11645A...
  • Page 33: Replaceable Parts

    Replaceable Parts 11645A...
  • Page 34: Introduction

    To order a listed part, note the description, the part number, and the quantity desired. Telephone or send your order to Agilent Technologies. Refer to “Contacting Agilent” on page 5-5 for further information.
  • Page 35 Replaceable Parts Introduction Figure 6-1 R-Band Component Identification Sheet 11645A...
  • Page 36 Replaceable Parts Introduction Table 6-2 Replaceable Parts for the Q11645A WR-22 Description Agilent Per Kit Part Number Attenuators 20 dB attenuator with data 11645-60023 50 dB attenuator with data 11645-60024 Airlines 50Ω airline with data 11645-60017 25Ω mismatch airline with data 11645-60012 Miscellaneous Items Disk holder...
  • Page 37 Replaceable Parts Introduction Figure 6-2 Q-Band Component Identification Sheet 11645A...
  • Page 38 Replaceable Parts Introduction Table 6-3 Replaceable Parts for the U11645A WR-19 Description Agilent Per Kit Part Number Attenuators 20 dB attenuator with data 11645-60025 50 dB attenuator with data 11645-60006 Airlines 50Ω airline with data 11645-60018 25Ω mismatch airline with data 11645-60013 Miscellaneous Items Disk holder...
  • Page 39 Replaceable Parts Introduction Figure 6-3 U-Band Component Identification Sheet 11645A...
  • Page 40 Replaceable Parts Introduction Table 6-4 Replaceable Parts for the V11645A WR-15 Description Agilent Per Kit Part Number Attenuators 20 dB attenuator with data 11645-60007 50 dB attenuator with data 11645-60008 Airlines 50Ω airline with data 11645-60019 25Ω mismatch airline with data 11645-60014 Miscellaneous Items Disk Holder...
  • Page 41 Replaceable Parts Introduction Figure 6-4 V-Band Component Identification Sheet 11645A...
  • Page 42 Replaceable Parts Introduction Table 6-5 Replaceable Parts for the W11645A WR-10 Description Agilent Per Kit Part Number Attenuators 20 dB attenuator with data 11645-60009 50 dB attenuator with data 11645-60010 Airlines 50Ω airline with data 11645-60020 25Ω mismatch airline with data 11645-60015 Miscellaneous Items Disk holder...
  • Page 43 Replaceable Parts Introduction Figure 6-5 W-Band Component Identification Sheet 11645A 6-11...
  • Page 44 Replaceable Parts Introduction Table 6-6 Items Not Included in the Verification Kit Description Agilent Part Number ESD Protection Devices Grounding wrist strap 9300-1367 5 ft grounding cord for wrist strap 9300-0980 2 x 4 ft conductive table mat and 15 ft ground wire 9300-0797 ESD heel strap (for conductive floors) 9300-1308...
  • Page 45 Index Agilent Technologies damage handling contacting device heel strap part number airline, part numbers inspecting for 6-12 humidity, specifications 6-10 what to do alcohol, isopropyl data, recertification precautions for use of device altitude, specifications cleaning incoming inspection assistance compatible analyzers...
  • Page 46 Index serial, recording supplies cleaning part numbers 6-12 ordering parts oxygen tag, service temperature affect on electrical performance part numbers error-corrected included in kit 6-10 operating not included in kit 6-12 specifications performance test what to watch out for performance verification, failure test data preventive maintenance test, performance...

This manual is also suitable for:

R11645aQ11645aU11645aV11645aW11645a