7
Performance Tests
List of Test Equipment
For scope performance verification, the recommended test equipments are
listed as below.
Table 7-1 List of test equipments for scope performance test
Application
Voltage measurement
accuracy
Bandwidth
Horizontal time base
accuracy
Trigger sensitivity
Agilent U1602B/U1604B User's and Service Guide
Equipment
Calibrator / Power supply
Digital multimeter
Cable
Shorting capacitor
Adapter
Adapter
Calibrator
Feedthrough
Cable
Calibrator
Feedthrough
Cable
Calibrator
Power splitter
Cable
Adapter
Feedthrough
Critical Specifications
5 mV to 35 VDC, 0.1 mV resolution
Better than 0.01% accuracy
BNC, quantity of 2
BNC
BNC (f) to banana (m)
BNC tee (m) (f) (f)
NA
50&, BNC connector (m) and (f)
Type N (m), 24 inches
5 ppm after 1/2 an hour
50&, BNC connector (m) and (f)
BNC, 3 feet length
25 MHz, 60 MHz and 100 MHz sine
waves.
Outputs differ < 0.15 dB
BNC, quantity of 3
N (m) to BNC (f), quantity of 3
50&, BNC connector (m) and (f),
quantity of 2
Recommended Equipment
FLUKE 5520A
Agilent 34401A
Agilent 10503A
Agilent 1250-0774
Agilent 1251-2277
Agilent 1250-0781
FLUKE 5520A
Agilent 11048C
Agilent 11500B
FLUKE 5520A
Agilent 10503A
Agilent 11048C
FLUKE 5520A
Agilent 11667B
Agilent 10503A
Agilent 1250-0780
Agilent 11048C
133