Performance Tests

Continuity Verification Test

1 Put the calibrator in Standby (STBY) mode.
2 Select Meter mode of the instrument, then press F2 softkey for
3 Press F1 to select continuity test.
4 Connect the calibrator to the instrument's input and ground terminal as
5 Apply a resistance output of 30
6 Apply a resistance output of 75
Agilent U1602B/U1604B User's and Service Guide
Ohmmeter.
shown in Figure 7- 2 using banana plug.
sound.
sound.
Ω
from the calibrator, buzzer should
Ω
from the calibrator, buzzer should not
149