Dc Volts Verification Test - Agilent Technologies U1602B User's And Service Manual

Handheld digital oscilloscopes
Hide thumbs Also See for U1602B:
Table of Contents

Advertisement

Table 7-5 DC Volts Verification Test
Calibrator Voltage
600 mV
6 V
30 V
60 V
600 V
Figure 7-2 Test connections from calibrator to instrument
144
Function
Range
DC Volts
600 mV
6 V
60 V
60 V
600 V
Min Limit
597.7 mV
5.977 V
29.86 V
59.77 V
597.7 V
Agilent U1602B/U1604B User's and Service Guide
Performance Tests
Max Limit
602.3 mV
6.023 V
30.14 V
60.23 V
602.3 V
7

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1604bU1602aU1604a