Performance Tests

Capacitance Verification Test

1 Put the calibrator in Standby (STBY) mode.
2 Select Meter mode of the instrument, press F2 softkey for Ohmmeter.
3 Press F1 to select capacitance measurement.
4 Connect the calibrator to the instrument's input and ground terminal as
5 Select each function and range in the order shown in Table 7- 8. Apply
6 Compare the measurement results to the corresponding test limits
Table 7-8 Capacitance Verification Test
Calibrator Settings
60 nF
600 nF
6 mF
36 mF
60 mF
300 mF
Agilent U1602B/U1604B User's and Service Guide
shown in Figure 7- 2 using a banana plug.
the values from the calibrator shown in table 7- 8 below.
shown in table 7- 8.
Function
Range
Capacitance
60 nF
600 nF
6 mF
60 mF
60 mF
300 mF
Min Limit
Max Limit
58.68 nF
61.32 nF
586.8 nF
613.2 nF
5.868 mF
6.132 mF
35.16 mF
36.84 mF
58.68 mF
61.32 mF
293.4 mF
306.6 mF
147