Performance Tests
4 Press Autoscale on the instrument.
5 Select Measure mode, then press F1 to enable the auto measurement
6 View the Pk- Pk reading at the measurement line display. Record the
7 Change the frequency of the signal generator to the value shown below
Table 7-4 Signal generator frequency setting
8 Change the time base to 50 ns/div.
9 Wait for a few seconds for the measurement to settle.
10 View the Pk- Pk reading at the measurement line display.
11 Record the reading: Vp- p = ______ mV.
12 Calculate the bandwidth response (dB) using the following formula:
13 The result from step 12 should be greater than or equal to –3.0 dB.
14 Move cable from channel 1 to channel 2 input and repeat the same
Agilent U1602B/U1604B User's and Service Guide
selection for Measure 1. Turn rotary switch to select peak- to- peak
voltage (Pk- Pk) measurement from the list of auto measurement.
Wait for a few seconds for the measurement to settle.
reading: Vp- p = _______ V. (refer to
Digital Oscilloscope Test
for your instrument.
Selected Channel
Channel 1
Channel 2
20log
[Reading from step #11 / Reading from step #6]
10
verification procedures for channel 2.
"Agilent U1602B/U1604B Handheld
Record")
U1602B
20 MHz
20 MHz
U1604B
40 MHz
40 MHz