Section 4 Initial Setup

4.2 Initial Setup Example for the K3HB-P

4-4
The initial setup is explained in the following example.
Settings Example
The passing speed is displayed in m/s when the distance between A
and B is 5 m.
• If the measurement value goes above 0.700, comparative output
H turns ON.
• If the measurement value goes below 0.500, comparative output L
turns ON.
Measurement
value
0.700
0.500
Comparative output H
Comparative output L
Setting the Prescale Value
The prescale value can be obtained using the following formula when
the output is to be displayed in m/s.
Prescale value ( ) = 5/60 = 0.08333 ... = 8.3333
Prescale value of Input A, X (mantissa): ps. ax = 8. 3333
Prescale value of Input B, Y (exponent): ps. ay = 10 -2
Initial Setup Flow
●To change a set value, press the S [SHIFT] Key once to enable
changing the setting and then press the U [UP] Key to change the
value.
Press the M [MODE] Key to register the set value. The set value will
be registered and the next parameter will be displayed.
A Check the wiring and turn the power ON.
• The display will show "-----".
B Set the function to F1 (passing speed).
1. Move to the initial setting level by pressing the L [LEVEL] Key for at least
3 s (operation will stop).
2. Set "func" to "f1" and press the
C Set input type A and input type B to 00 (no-contact, normally
open).
1. Set input type A "in-ta" to "00" and press the
2. Set input type B "in-ta" to "00" and press the
A
B
2
10
M
[MODE] Key.
M
[MODE] Key.
M
[MODE] Key.