■ Characteristics
K3HB-R
Display range
19,999 to 99,999
Measurement accuracy
Functions F1, F6:
(at 23±5 C)
Functions F2 to F5: ±0.02% rdg ±1 digit (for voltage pulse/open collector sensors)
Measurement range
Functions F1 to F6: 0.5 mHz to 50 kHz (for voltage pulse/open-collector sensors)
Input signals
No-voltage contact
Voltage pulse
Open collector
Connectable sensors
ON residual voltage: 3 V max.
OFF leakage current: 1.5 mA max.
Load current:
Comparative output re-
Functions F1 to F6: 100 ms max. (time until the comparative output is made when there is a forced
sponse time (transistor
sudden change in the input signal from 15% to 95% or 95% to 15%.)
output)
Linear output response
Functions F1 to F6: 110 ms max. (time until the final analog output value is reached when there is a
time
forced sudden change in the input signal from 15% to 95% or 95% to 15%.)
Insulation resistance
20 M min. (at 500 VDC)
Dielectric strength
2,300 VAC for 1 min between external terminals and case
Noise immunity
100 to 240 VAC models:
24 VAC/VDC models:
Vibration resistance
Frequency: 10 to 55 Hz; Acceleration: 50 m/s
Shock resistance
150 m/s
Weight
Approx. 300 g (Base Unit only)
Degree of
Front panel
Conforms to NEMA 4X for indoor use (equivalent to IP66)
protection
Rear case
IP20
Terminals
IP00 + finger protection (VDE0106/100)
Memory protection
EEPROM (non-volatile memory)
Number of rewrites: 100,000
Applicable standards
UL61010C-1, CSA C22.2 No. 1010.1 (evaluated by UL)
EN61010-1 (IEC61010-1): Pollution degree 2/Overvoltage category II
EN61326: 1997, A1: 1998, A2: 2001
EMC
EMI: EN61326+A1 industrial applications
Electromagnetic radiation interference
Terminal interference voltage
EMS: EN61326+A1 industrial applications
Electrostatic Discharge Immunity
Radiated Electromagnetic Field Immunity
Electrical Fast Transient/Burst Immunity
Surge Immunity
Conducted Disturbance Immunity
Power Frequency Magnetic Immunity
Voltage Dips and Interruptions Immunity
±0.006% rdg ±1 digit (for voltage pulse/open collector sensors)
(30-Hz max. with ON/OFF pulse width of 15 ms min.)
(50-KHz max. with ON/OFF pulse width of 9 µs min.; ON voltage: 4.5 to 30 V;
OFF voltage: 30 to 2 V; input impedance: 10 k )
(50-KHz max. with ON/OFF pulse width of 9 µs min.)
Must have a switching capacity of 20 mA or higher.
Must be able to properly switch load currents of 5 mA or less.
1,500 V at power supply terminals in normal or common mode
(waveform with 1-ns rising edge and pulse width of 1 s/100 ns)
1,500 V at power supply terminals in normal or common mode
(waveform with 1-ns rising edge and pulse width of 1 s/100 ns)
2
2
(100 m/s
for relay outputs) 3 times each in 3 axes, 6 directions
CISPR 11 Group 1, Class A: CISPRL16-1/-2
CISPR 11 Group 1, Class A: CISPRL16-1/-2
EN61000-4-2: 4 kV (contact), 8 kV (in air)
EN61000-4-3: 10 V/m 1 kHz sine wave amplitude modulation (80 MHz to 1 GHz, 1.4 to 2 GHz)
EN61000-4-4: 2 kV (power line), 1 kV (I/O signal line)
EN61000-4-5: 1 kV with line (power line), 2 kV with ground (power line)
EN61000-4-6: 3 V (0.15 to 80 MHz)
EN61000-4-8: 30 A/m (50 Hz) continuous time
EN61000-4-11: 0.5 cycle, 0 /180 100% (rated voltage)
2
, 10 sweeps of 5 min each in X, Y, and Z directions
Specifications
A-3