5.15 Holding Comparative Outputs
f
o-stp
L
(O-STP)
0
L
Displays "L 0."
amov
0
L
0
amov
0
L
00000
The comparative output hold function holds the status of all outputs
after any output except for the PASS output turns ON, i.e., it stops
refreshing outputs. You can choose to stop outputs and continue
measurement, or to stop both.
Outputs will be refreshed again after the reset operation.
• "5.7 Resetting Measurements"
● Example with Output Refresh Stop ON
Comparative
set value HH
Comparative
set value H
Comparative
set value L
RESET
Output HH
Output H
Output L
Parameter
Output refresh stop
o-stp
Parameter Setting Procedure
A Press the L [LEVEL] Key for at least 3 s in RUN level to move to
L
the initial setting level.
3 s min.
0" is displayed on the level/bank display to indicate the initial
• "
L
setting level.
B Press the M [MODE] Key several times to change the PV display
M
to "amov."
• This parameter is not displayed for the initial status due to setting
level protect.
Refer to "5.34 Limiting Key Operations" (P.5-84) for information on
removing setting level protect.
C Press the S [SHIFT] Key to make the SV display flash.
S
• The setting can be changed when the SV display starts to
flash.
5.15 Holding Comparative Outputs
Advanced function setting level
P.5-33
Does not turn ON.
Meaning of set value
Set value
Outputs
off
Continue
out
Stop
all
Stop
R
P
Measurement
Continue
Continue
Stop
5-49