Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
2 Room Layout Drawings
LOCATION IN TECHNICAL ROOM FOR ELECTRICAL CABINETS IS MANDATORY.
THE ELECTRONIC CABINETS (C1, C2, COOLIX 4000 CHILLER, DETECTOR
CHILLER AND FLUORO UPS WHEN INSTALLED ) INCLUDE FANS THAT ARE
CREATING AIR–CIRCULATION OF PULSED–AIR. WHEN THIS PULSED AIR IS
IN AN ENVIRONMENT THAT MAY CONTAIN AIRBORNE PATHOGENS LIKE AN
EXAM ROOM/CONTROL ROOM, THERE IS A RISK OF TRANSMISSION OF THESE
AIRBORNE PATHOGENS FROM PATIENTS TO OTHER PATIENTS OR CLINICAL
PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE INSTALLED IN A
ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM, I.E., TECHNICAL ROOM.
For exam and control rooms, several configurations are recommended.
Motion controls installed in remote location from the table shall be installed at a location
NOTE:
where all the positioner axis are visible by the operator.
In case the system has a tilting table, the remote motion controls shall not be installed
on the longitudinal axis of the Table (to avoid any operator visual dead angle due to tilted
table top hiding the patient).
Refer to
Section 2.2
Chapter 4 Room Layouts
to
Section 2.6
to see recommended exam / control rooms layouts.
Direction 5314528-2-1EN, Revision 1
81