IV apparatus.
Injectors.
Heart monitoring equipment.
Crash cart.

2.3.5 Emergency Stop

Protect the Emergency Stop from accidental actuation.

2.3.6 Patient Environment Equipment

The components that may be installed within patient vicinity need to be medical equipment
("patient vicinity is defined in the standardization as a space within the room 1.83 m (70.7 in)
beyond the perimeter of the examination table and extending vertically 2.29 m (90.2 in) above
the floor."). For the Optima System, the equipment are:
Table
C-arm
Monitors
Injector
Rad-Shield
Patient Interface Module(s).
Table Side User Interfaces (TSUI)
Smart Box
Table Side System Control
Table Panning Handle
ITU
NOTE:
Central touch screen
Accessories:
Table Head Extender
Armboard
IV pole
Quick Strap
Head holder: Attach Head Positioning device onto the head end of the table with the
head cradle pointing to the end of the table. The clamp angle pressure bar should be
100
No IPX4 components shall be installed on a cart.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual