Hole Dimension And Preferred Location In Concrete Floor - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Prepare the floor such that the Tilt Table base plate will be flush with the floor finish
surface, taking into account the thickness of the floor finish material.
For alternative table bolts or seismic area, refer to template drawing shown in Illustration
Innova LC Positioner And Table Floor Mounting Template contained in
Mounting Requirements.

3.3.4 Hole dimension and preferred location in concrete floor

In the examination room, the Innova LC Positioner is not placed on a computer floor but
directly put on concrete floor, the location of the cable access needs to be carefully planned.
Otherwise, if the cable run is located under the concrete floor, the cables will have to come
through the floor and in this case you will need two holes, one for the LC Positioner and the
other for the patient table.
The diameter of both holes must be the same 225 mm or 9 in.
Chapter 3 Physical Requirements of Room
Direction 5314528-2-1EN, Revision 1
Chapter 5, Section 2,
61

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3100Innova 3100-iqInnova 4100Innova 4100-iq

Table of Contents