Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare

5.3 Run Interconnection cables through the Conduit

Use the IVUS PIM (#806365003) as reference to the procedures below.
Unpack , and check the completeness of the Preinstallation kit :
Install 806889001 Ground Cable through the Control Booth conduit (see note)
Install the cables through the IVUS conduit from the Control Room end :
Install the optional cables through the IVUS conduit from the Control Room end :
NOTE:
For the ground cable, see the IVUS service manual for special instructions concerning
cable running and cutting .
NOTE:
If the installation has the MaC-Lab ECG in the configuration, the ECG cable shall not be run
through the conduit, it will be used in the Control Room to connect Mac-Lab and IVUS PCs.
NOTE:
You should consider running the ECG cable through the conduit in case, the customer
requires interconnecting to a NON Mac-Lab ECG at the table base.
In this case you shall hide the unconnected cable in the conduit for future expansion
options.
NOTE:
Same applies to the 2nd Ethernet cable, if a later extension of the IVUS remote controls
expected.
Ensure sufficient cable length on the Procedures Room end of the cables
according to the table below
PIM cable
Control Station cable
Joystick cable
ECG cable
The incorrect cable length might cause reliability issues during the operation !
NOTE:
Make sure, that the out of conduit parts of the cables with the connectors are properly
protected during the rest of the installation
170
1pcs - Custom PIM cable 30 m (98 ft, 5 in) - Volcano ID :806452001
2pcs - CAT5 cable 30 m (98 ft, 5 in) - Volcano ID :803391001
1pcs - Custom Twin axial cable - Volcano ID :806380001
1pcs – Ground Cable, 30 m (98 ft, 5 in) s5i – Volcano ID :806889001
1pcs - IVUS preinstallation Manual - Volcano ID :806365003
806452001 Room - PIM cable, round shaped connector end to the Procedures
803391001 - Ethernet cable (1pcs)
803391001 - Ethernet cable if 2 IVUS remote control items will be installed 1 IVUS
Control Station and 1 IVUS Joystick)
806380001 - ECG cable in case, if NON Mac-Lab ECG will be installed
Cable
Direction 5314528-2-1EN, Revision 1
Free length from the conduit exit
2.4 m (8 ')
2.1 m (7 ')
2.7 m (9 ' )
Depends on the ECG location
5 IVUS Option