Important Notice Concerning Power Cable Selection - Siemens RM400 C Series Operating Manual

Table of Contents

Advertisement

2
Important notes
2.5
Important notice concerning power cable
selection
The power cable for this unit has been packed separately. It has been selected
according to the country of destination and must be used to prevent electric shock.
Use the following guidelines to replace the original cable set if necessary. The
female receptacle of the cable set must meet CEE-22 requirements.
Figure 1: Female connector
For the USA and Canada
Use a UL-listed and CSA-labeled cable set consisting of a three-conductor cable
with a maximum length of 15 feet.
For desktop units, type SVT or SJT cable sets must be used. For floor units, only
SJT type cable sets can be used.
The cable set must be selected according to the current rating for your device.
Please consult the table below for the selection criteria for power cables used in the
USA or Canada.
Cable type
Size of conductors in cable
SJT
18 AWG
16 AWG
14 AWG
SVT
18 AWG
17 AWG
12
Maximum current rating of
device
10 amps
12 amps
12 amps
10 amps
12 amps
RM400 C Models
U24802-J-Z146-2-7600