Preparation for Measurement

Initialize the System

Initialize the System
Before placing the HGA cassette on the HLM of the spinstand, initialize the system. In the
case of VEE measurement program, start the program before placing the HGA cassette.
This is because the initialization perform the homing of the spinstand. As the homing
moves the stage to full travel. This might that the HGA contacts the shroud.

Setting the HGA Cassette

Set the head cassette on the HLM (Head Loading Mechanism). Cassette attachment is
basically the same for all types of cassettes. Push the cassette knob inwards to widen its
claw, then insert the head cassette on top of the HLM. Release the cassette knob and check
if it is locked down on the HLM. The cassette face (i.e, up face, down face) is
automatically detected by the cassette sensor mounted on each side of the HLM.
Figure 2-4
Setting the Cassette on the HLM (E5010B)