Cassette and Z-Height Adjustment

Z-height Adjustment for E5029C/D Cassettes

Z-height Adjustment for E5029C/D Cassettes
The Z-height adjustment tool is used to adjust the Z-height of the HGA mount face. The
Z-height Adjustment tool is included in the Agilent E5029C Option Zxx (xx represents a
two-digit number). This section provides the Z-height adjustment procedure of E5029C
Option C11. Other cassettes have the same adjustment procedure.
Table C-1
1. Prepare the Z-height Adjustment tool. The Z-height Adjustment tool is included in the
Figure C-1
Z-height adjustment tool
2. Select one of the three pin holes that will be used to measure the cassette's Z-height as
NOTE
The location of the needle pin of the Z-height tool (i.e. short, medium or long) should
match the length of the HGA Mount block assembly in which the user decides to use. Refer
to Figure C-12.
392
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Required Tools
Tools
Hex Screwdriver
Hex Screwdriver
Hex Screwdriver
Slotted Screwdriver
Agilent E5029C Option Zxx (xx represents a two-digit number).
shown in Figure C-2 and Figure C-3.
Size
1.5 mm
0.89 mm
0.71 mm
Appendix C