Service Release For Revision A.04.10; Changes At Revision A.04.10.03; Changes At Revision A.04.10.02 - Agilent Technologies E5035A Operation Manual

Hard disk read/write test system
Table of Contents

Advertisement

Manual Changes

Service Release for Revision A.04.10

Service Release for Revision A.04.10

Changes at Revision A.04.10.03

1. Supports the following Head Amplifiers
E5029J Opt.613 E5022-66613/E5022-66614 VM8020
2. Supports the following Cassette
E5029D Opt.009 E5010-61291/E5010-61292

Changes at Revision A.04.10.02

1. Supports the following Head Amplifiers
E5029J Opt.107 E5022-66107/E5022-66108 SR1759-2
Changes at Revision A.04.10.02
1. The following bug is fixed
Symptom: When click on [Head Amp Config] under [Setup][Configuration] on the
VEE program, only installed head amplifier's part number and its serial number are
displayed and you can not set the control parameters, such as gain, bias and
overshoot. This is happen on head amplifiers which has a property function except
E5029J Option 609 (E5022-66609/E5022-66610). This is a bug in the VEE
Program and the functions are no problem.