Zero Offset Adjustment - Agilent Technologies 16334A Operation And Service Manual

Test fixture

Advertisement

Operation

ZERO OFFSET ADJUSTMENT

ZERO OFFSET ADJUSTMENT
The 16334A has inherent residual impedance and stray admittance that affects the accuracy
of the measurements. To compensate for, or negate, these residuals in order to minimize
measurement error, the instrument's zero offset adjustment procedure should be performed.
The zero offset adjustment procedure is described in the instrument's manua1. When
performing the zero offset adjustment, use compensation block Agilent PN 16334-60001
(furnished with probe). An illustration showing how to use the compensation block is
shown in Figure 1-3.
Figure 1-3
Zero Offset Adjustment
14
Chapter 1