Fluke 96000 Series Service Manual page 95

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

Connect the Signal Source
and Spectrum Ananlyzer Frequency
Reference Inputs to the UUT
Reference Output
Instrument (UUT)
75 Ω Leveling
Figure 3-14. Equipment Connections - 75 Ω Levelling Head VSWR Test (96040A and 96270A)
12. Compute the average of the values measured with the test port open and
shorted. Record this number as Z
96xx
HEAD INTERFACE
RF OUTPUT
CONTROL
CONNECT
96xx HEAD
ONLY
Spectrum Analyzer
75 Ω to 50 Ω
Impedance
Matching Pad
Head
Open
Short
for the test frequency of 500 MHz.
Max
Signal Source
9610A/AF REFERENCE SOURCE
SINE
MOD
SWEEP
UNITS
10V
PK
MAX
SETUP
RF
COMMON
RF Input
75 Ω Leveling
Input Port
75 Ω Directional Bridge
Test Port
3