96000 Series
Service Manual
High Frequency Test Points (steps 8-17)
8. Connect the 75 Ω Leveling Head to the 75 Ω Power Sensor via a precision
Figure 3-11. Equipment Connections- 75 Ω Leveling Head, High Frequency Level Accuracy Test
9. Set the UUT as follows:
10. Configure the Power Meter to indicate readings in dBm. Enter the
3-40
75 Ω Type-N female-to-female adapter. See Figure 3-11 for equipment
connections.
Instrument
75
Points (96040A and 96270A)
Mode
Frequency
Level
Output
measurement frequency (0.1 MHz) into the Power Meter to enable cal factor
(frequency) correction.
N Adapter
Head
Leveling
Leveled Sine
100 kHz
+10 dBm
OPER
Power Meter
A
B
Power Sensor
hpn215.eps