Harmonics and Spurious Signal Content – Levelling Head (96040A and 96270A)
Use a high performance spectrum analyzer to make the measurements shown in
Tables A-3 and A-4.
Output
Level
Frequency
20 kHz
2.74 MHz
5.5 MHz
11 MHz
+24 dBm
22 MHz
44 MHz
62.5 MHz
88 MHz
125 MHz
250 MHz
354 MHz
500 MHz
+20 dBm
714 MHz
1 GHz
1.4 GHz
1.8 GHz
+14 dBm
2.7 GHz
4.024 GHz
Table A-3. Harmonics Test
nd
2
Harmonic
Freq
Limit
40 kHz
-60 dBc
5 MHz
-60 dBc
11 MHz
-60 dBc
22 MHz
-60 dBc
44 MHz
-60 dBc
88 MHz
-60 dBc
125 MHz
-60 dBc
176 MHz
-60 dBc
250 MHz
-60 dBc
500 MHz
-60 dBc
708 MHz
-60 dBc
1 GHz
-60 dBc
1.428 GHz
-60 dBc
2 GHz
-60 dBc
2.8 GHz
-55 dBc
3.6 GHz
-55 dBc
5.4 GHz
-55 dBc
8 GHz
-55 dBc
rd
3
Harmonic
Actual
Freq
60 kHz
-60 dBc
7.5 MHz
-60 dBc
16.5 MHz
-60 dBc
33 MHz
-60 dBc
66 MHz
-60 dBc
132 MHz
-60 dBc
187.5 MHz
-60 dBc
264 MHz
-60 dBc
375 MHz
-60 dBc
750 MHz
-60 dBc
1.062 GHz
-60 dBc
1.5 GHz
-60 dBc
2.142 GHz
-60 dBc
3 GHz
-60 dBc
4.2 GHz
-55 dBc
5.4 GHz
-55 dBc
8.1 GHz
-55 dBc
12 GHz
-55 dBc
Performance Test Record
Performance Test Record
Limit
Actual