Frequency Accuracy – Microwave Output (96270A)
Use a high resolution frequency counter, spectrum analyzer and external
frequency reference to make the measurements shown in Table A-23.
10 MHz
1 GHz
5 GHz
6.036 GHz
8.048 GHz
9.52 GHz
10 GHz
11.32 GHz
12.072 GHz
13.5 GHz
15 GHz
16 GHz
18.108 GHz
19.04 GHz
20 GHz
21.84 GHz
22.64 GHz
24 GHz
26.5 GHz
A-24
Table A-23. Frequency Accuracy Test
Frequency
Actual
(After 24 hour warm-up)
±0.5 Hz
±50 Hz
±250 Hz
±301.8 Hz
±402.4 Hz
±476 Hz
±500 Hz
±566 Hz
±603.6 Hz
±200.0 Hz
±750 Hz
±800 Hz
±905.4 Hz
±952 Hz
±1 kHz
±1.090 kHz
±1.132 kHz
±1.2 kHz
±1.325 kHz
Tolerance