Fluke 96000 Series Service Manual page 49

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

4. Set the UUT as follows:
Ref Frequency Input
Mode
Signal
Frequency
Level
Output
5. Set the Frequency Counter as follows:
Function
Slope
Coupling
Input Impedance
Attenuation
Trigger Mode
6. Set the output frequency of the UUT to each of the frequencies listed in
Part A of Table 3-2. At each frequency, let the Frequency Counter take
several readings and then settle. The settled reading must be within the
tolerance shown. For frequencies of ≥500 MHz connect the UUT Leveling
Head to the counter channel C input and select the counter Freq C function.
7. Set the UUT output to STBY. Remove all connections to the Frequency
Counter.
Part
A
B
8. Set the UUT output to STBY. Remove all connections to the Frequency
Counter.
9. Connect the 10 MHz output from the Frequency Standard to the EXT REF IN
on the Spectrum Analyzer (rear panel) using a BNC male – male cable
assembly. Set the Spectrum Analyzer for external frequency reference.
10. Connect the UUT Microwave output to the input of the Spectrum Analyzer.
See Figure 3-3 for equipment connections.
Disabled
Leveled Sine
Microwave Output
10 kHz
+13 dBm
OPER
Freq A
Positive
AC
50 Ω
1x
Auto
Table 3-2. Frequency Accuracy Test
UUT Frequency
10 kHz
9.999 999 495 – 10.000 000 505 kHz
10 MHz
9.999 999 500 – 10.000 000 500 MHz
30 MHz
29.999 998 500 – 30.000 001 500 MHz
50 MHz
49.999 997 500 – 50.000 002 500 MHz
100 MHz
99.999 995 000 – 100.000 005 000 MHz
500 MHz
499.999 980 – 500.000 020 MHz
1 GHz
0.999 999 950 0 – 1.000 000 050 0 GHz
2 GHz
1.999 999 900 0 – 2.000 000 100 0 GHz
2.7 GHz
2.699 999 865 0 – 2.700 000 135 0 GHz
4 GHz
3.999 999 800 0 – 4.000 000 200 0 GHz
Performance Tests
Performance Test
Tolerance
3
3-7