Table 12. Level Accuracy (50 Ω) Test (cont).
Level
1 kHz
20 kHz
100 kHz
10 MHz
-37 dBm
100 MHz
300 MHz
1.4 GHz
3 GHz
4 GHz
1 kHz
20 kHz
100 kHz
10 MHz
-47 dBm
100 MHz
300 MHz
1.4 GHz
3 GHz
4 GHz
100 kHz
10 MHz
100 MHz
-57 dBm
300 MHz
1.4 GHz
3 GHz
4 GHz
100 kHz
10 MHz
100 MHz
-66dBm
300 MHz
1.4 GHz
3 GHz
4 GHz
Frequency
Performance Test Record
Performance Test Record
Actual
Tolerance
±0.03 dB
±0.03 dB
±0.03 dB
±0.05 dB
±0.05 dB
±0.07 dB
±0.2 dB
±0.3 dB
±0.5 dB
±0.03 dB
±0.03 dB
±0.03 dB
±0.05 dB
±0.05 dB
±0.07 dB
±0.2 dB
±0.3 dB
±0.5 dB
±0.2 dB
±0.1 dB
±0.1 dB
±0.1 dB
±0.4 dB
±0.5 dB
±0.5 dB
±0.2 dB
±0.1 dB
±0.1 dB
±0.1 dB
±0.4 dB
±0.5 dB
±0.5 dB