Frequency Accuracy (96040A and 96270A)
Use a high resolution frequency counter, spectrum analyzer and external
frequency reference to make the measurements shown in Table A-2.
10 kHz
10 MHz
30 MHz
50 MHz
100 MHz
500 MHz
1 GHz
2 GHz
3 GHz
4 GHz
A-2
Table A-2. Frequency Accuracy Test
Frequency
Actual
(After 24 hour warm-up)
±0.505 mHz
±0.5 Hz
±1.5 Hz
±2.5 Hz
±5.0 Hz
±25.0 Hz
±50.0 Hz
±100.0 Hz
±135.0 Hz
±200.0 Hz
Tolerance