Fluke 96000 Series Service Manual page 174

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

96000 Series
Users Manual
Phase Noise (96040A and 96270A, Optional)
Use a signal source analyzer to make the measurements shown in Table A-5.
Output
125 MHz
1 GHz
A-4
Table A-4. Spurious Signal Content Test
Frequency
2.150 003 GHz
2.275 003 GHz
2.425 003 GHz
2.500 003 GHz
2.650 003 GHz
2.825 003 GHz
2.999 997 GHz
3.149 997 GHz
3.424 997 GHz
3.499 997 GHz
3.649 997 GHz
3.774 997 GHz
3.924 997 GHz
3.999 997 GHz
Table A-5. Phase Noise Test
Offset
Actual
100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
10 MHz
100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
10 MHz
Spurious Level
< -66 dBc
< -66 dBc
< -66 dBc
< -66 dBc
< -66 dBc
< -66 dBc
< -66 dBc
< -66 dBc
< -66 dBc
< -66 dBc
< -66 dBc
< -66 dBc
< -66 dBc
< -66 dBc
Specification
Uncertainty
< -121 dBc/Hz
< -141 dBc/Hz
< -148 dBc/Hz
< -151 dBc/Hz
< -153 dBc/Hz
< -155 dBc/Hz
< -104 dBc/Hz
< -130 dBc/Hz
< -140 dBc/Hz
< -141 dBc/Hz
< -148 dBc/Hz
< -152 dBc/Hz
Limit
Limit