Fluke 96000 Series Service Manual page 100

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

96000 Series
Service Manual
4. Set the UUT output to the next frequency listed in Table 3-24 for this test
5. Repeat step 3 and confirm that the measured output level is consistent with
6. When the test frequency is 100 kHz, record the measured level in dBm as P
7. Repeat steps 2 through 7 for the next test amplitude listed in Table 3-24 for
Table 3-24. Microwave Output Level Accuracy Test, Low Frequency Test Points
Low Level Microwave Output
All UUT configurations
Non-Low Level MW O/P Option
only
Low Level Microwave Output
Option only
3-58
amplitude.
the typical performance shown in Table 3-24.
for use later in this procedure.
the relevant UUT Low Level Microwave Output option configuration, applying
the 5790A 2.2 mV range linearity error correction at levels below -40 dBm.
Option?
+13 dBm
+3 dBm
-4 dBm
-7 dBm
-17 dBm
-27 dBm
-37 dBm
-47 dBm
Amplitude
Frequency
1 kHz
100 kHz
1 kHz
100 kHz
1 kHz
100 kHz
1 kHz
100 kHz
1 kHz
100 kHz
1 kHz
100 kHz
1 kHz
100 kHz
1 kHz
100 kHz
Typical Tolerance
±0.5 dB
(P
)
±0.5 dB
1
±0.5 dB
(P
)
±0.5 dB
1
±0.5 dB
(P
)
±0.5 dB
1
±2.0 dB
±2.0 dB
±2.0 dB
±2.0 dB
±2.0 dB
(P
)
±2.0 dB
1
±2.0 dB
(P
)
±2.0 dB
1
±2.0 dB
(P
)
±2.0 dB
1
1