11. Repeat steps 5 through 10 for all frequencies in Table 3-3, using the settings
and tolerances shown for each. For UUT output frequency settings of
1.8 GHz and above where the UUT level setting is +14 dBm, set the Analyzer
reference level to +20 dBm.
Level
Frequency
20 kHz
2.74 MHz
5.5 MHz
11 MHz
+24 dBm
22 MHz
44 MHz
62.5 MHz
88 MHz
125 MHz
250 MHz
354 MHz
500 MHz
+20 dBm
714 MHz
1 GHz
1.4 GHz
1.8 GHz
+14 dBm
2.7 GHz
4.024 GHz
12. Set the Spectrum Analyzer as follows:
PRESET
COUPLING
REF LVL
EXT REF
FREQ
SPAN
13. Set the UUT as follows:
Mode
Frequency
Level
Output
14. Set the Spectrum Analyzer to single sweep, initiate a sweep and wait for the
sweep to complete. Push MARKER → PEAK.
15. Using the Spectrum Analyzer marker delta functions, measure the amplitude
of any spurious signals detected. Verify any spurious signals at offsets
>±3 kHz from the output frequency within the listed tolerances.
Table 3-3. Harmonics Test
nd
2
Harmonic
Tolerance
40 kHz
-60 dBc
5.48 MHz
-60 dBc
11 MHz
-60 dBc
22 MHz
-60 dBc
44 MHz
-60 dBc
88 MHz
-60 dBc
125 MHz
-60dBc
176 MHz
-60 dBc
250 MHz
-60 dBc
500 MHz
-60 dBc
708 MHz
-60 dBc
1 GHz
-60 dBc
1.428 GHz
-60 dBc
2 GHz
-60 dBc
2.8 GHz
-55 dBc
3.6 GHz
-55 dBc
5.4 GHz
-55 dBc
8.048 GHz
-55 dBc
DC
+10 dBm
On
2.150 003 GHz
2 MHz
Leveled Sine
2.150 003 GHz
+4.6 dBm
OPER
Performance Tests
Performance Test
rd
3
Harmonic
Tolerance
60 kHz
-60 dBc
8.22 MHz
-60 dBc
16.5 MHz
-60 dBc
33 MHz
-60 dBc
66 MHz
-60 dBc
132 MHz
-60 dBc
187.5 MHz
-60dBc
264 MHz
-60 dBc
375 MHz
-60 dBc
750 MHz
-60 dBc
1.062 GHz
-60 dBc
1.5 GHz
-60 dBc
2.142 GHz
-60 dBc
3 GHz
-60 dBc
4.2 GHz
-55 dBc
5.4 GHz
-55 dBc
8.1 GHz
-55 dBc
12.072 GHz
-55 dBc
3
3-11