Fluke 96000 Series Service Manual page 188

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

96000 Series
Users Manual
Frequency
100 MHz
VSWR - 75
Use a directional bridge and a spectrum analyzer to make the measurements
shown in Table A-19.
+7 dBm
-3 dBm
-13 dBm
A-18
Table A-18. Attenuation Accuracy (75 Ω) Test (cont.)
Attenuation
0 dB
-3 dB
-13 dB
-23 dB
-33 dB
-43 dB
-53 dB
-63 dB
-73 dB
-82 dB
-91 dB
-101 dB
Ω
Levelling Head (96040A and 96270A, Optional)
Table A-19. VSWR (75 Ω) Test
Level
Frequency
500 MHz
1 GHz
2 GHz
500 MHz
1 GHz
2 GHz
500 MHz
1 GHz
2 GHz
Actual
Specification
---
Reference
±0.07 dB
±0.07 dB
±0.07 dB
±0.07 dB
±0.1 dB
±0.1 dB
±0.1 dB
±0.2 dB
±0.2 dB
±0.2 dB
±0.4 dB (Typ)
Actual
Uncertainty
Tolerance
Reference
Reference
Tolerance
<1.1
<1.2
<1.3
<1.1
<1.2
<1.3
<1.1
<1.2
<1.3

Advertisement

Table of Contents
loading

This manual is also suitable for:

96040a96270a

Table of Contents