Fluke 96000 Series Service Manual page 45

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

Equipment
Power Sensor, Diode, 50 Ω
Power Sensor, 75 Ω
Power Sensor, Thermal, 27 GHz, 50 Ω
[5], [6]
[7]
Common mode choke
, if required.
[1]
Required only for the optional phase noise test. Refer to the phase noise test section for further details of equipment selection.
[2]
Not required for 96040A.
If a 50 Ω Type-N feedthrough termination is unavailable, a Rohde & Schwarz RAD 50 (or equivalent) with a precision 50 Ω Type-N (f) to BNC (m)
[3]
adapter may be substituted.
If a 75 Ω Type-N feedthrough termination is unavailable, an Agilent 11094B (or equivalent) with a precision 75 Ω Type-N (f) to BNC (m) adapter may
[4]
be substituted.
[5]
Not required for 96040A. When used for the Microwave Output Level Accuracy test an equivalent power sensor and corresponding power meter
readout may be substituted. The model NRP-Z55.03 power sensor is supplied as part of the 96000 HF Levelling Kit and 96000SNS Additional Sensor
option.
[6]
Not required for 96040A. The model NRP-Z55.03 is supplied as part of the 96000 HF Levelling Kit and 96000SNS Additional Sensor option. When
used for the optional Microwave Output characterization adjustment only compatible Rhode & Schwarz NRP-Z series power sensors of the
appropriate frequency range may be substituted. Refer to chapters 1 and 3 of the 96000 Series Operator's Manual for details of compatible sensors.
No other substitutions are possible for this adjustment.
[7]
May be required for measurements made with the 5790A. Refer to relevant sections of this chapter and section 4-12 of the 5790A Operator's Manual
for further details.
Typical examples of the AC Measurement Standard listed in Table 3-1 may
exhibit a slight noise floor-related non-linearity on the lowest (2.2 mV) range
which can increase the uncertainty of the tests or adjustments unless a
correction is applied. To determine a 2.2 mV range linearity error, compare the
range with the linearity of the higher ranges where no significant non-linearity
exists.
Fluke Calibration recommends that users familiarize themselves with the
procedures documented in this chapter. The procedures indicate when these
Power Sensor and AC Measurement Standard characteristics must be taken into
account.
Users should also familiarize themselves with the detailed operation of the
equipment listed in Table 3-1 by reading the relevant manufacturer's instructions
and manuals.
Table 3-1. Recommended Equipment (cont.)
Recommended
Model
Rohde & Schwarz NRV-Z4
or
Rohde & Schwarz NRP-Z91
Rohde & Schwarz NRV-Z3
Rohde & Schwarz NRP-Z53
Fluke part number 474908
Recommended Tools and Equipment
Minimum Use Requirement
Type-N Connector
Frequency: 100 kHz to 4 GHz
Power: -10 to -70 dBm
Cal factor uncertainty: ±0.3%
Max. SWR: 0.1 to 100 MHz: 1.05
Type-N Connector
Frequency: 1 MHz to 2 GHz
Power: +10 to -20 dBm;
Cal factor uncertainty: ±0.3%;
Max. SWR: 1.15
2.92mm Connector
Frequency: 100 kHz to 26.5 GHz
Power: +20 to -25 dBm
Cal factor uncertainty: ±0.04dB
Max. SWR: 1.22; DC coupled
Approximately 250 µH, see text.
3
Performance Tests
3-3