Fluke 96000 Series Service Manual page 86

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

96000 Series
Service Manual
19. On the UUT enable the REF FREQUENCY OUTPUT a 10 MHz and set the
20. Allow the Measuring Receiver to tune the input signal, and then set the
21. Without changing any Measuring Receiver settings, set the UUT to the next
22. Calculate the UUT output P
23. Repeat steps 22 and 23 for each amplitude listed for this frequency in part A
24. Return the UUT output to -33 dBm, set the UUT frequency to the next
25. Set the UUT to 100 kHz. Repeat steps 21 through 25 for the points listed in
26. Set the UUT to STBY.
3-44
UUT as follows:
Frequency
Level
Output
controls to measure RF level, relative (dB) to the current input level.
amplitude for this frequency listed in Table 3-19. Allow the Measuring
Receiver reading to stabilize and record the reading as P
be a negative number in dB).
within the tolerance shown in part A of Table 3-19.
of Table 3-19.
frequency listed in part A of Table 3-19 and repeat steps 21 through 24.
part B of Table 3-19 using the calculation P
100 kHz
-33 dBm
OPER
= P
+ P
. Check that the value of P
out
-33
rel
= P
out
. (Note that P
rel
is
out
+ P
in step 22.
-43
rel
will
rel