Table 3-27. Microwave Output Level Accuracy Test, Additional Low Level Test Points, Part B
Part
Frequency
100 kHz
10 MHz
100 MHz
B
Low Level
Microwave
Output Option
only
1 GHz
4.024 GHz
10 GHz
Amplitude
-37 dBm
Reference
-57 dBm
P
rel
-66 dBm
P
rel
-75 dBm
P
rel
-85 dBm
P
rel
-37 dBm
Reference
-47 dBm
P
rel
-57 dBm
P
rel
-66 dBm
P
rel
-75 dBm
P
rel
-85 dBm
P
rel
-37 dBm
Reference
-47 dBm
P
rel
-57 dBm
P
rel
-66 dBm
P
rel
-75 dBm
P
rel
-85 dBm
P
rel
-37 dBm
Reference
-47 dBm
P
rel
-57 dBm
P
rel
-66 dBm
P
rel
-75 dBm
P
rel
-85 dBm
P
rel
-37 dBm
Reference
-47 dBm
P
rel
-57 dBm
P
rel
-66 dBm
P
rel
-75 dBm
P
rel
-85 dBm
P
rel
-37 dBm
Reference
-47 dBm
P
rel
-57 dBm
P
rel
-66 dBm
P
rel
Performance Tests
Reading
P
Tolerance
out
Reference
±2.0 dB
±2.0 dB
±2.0 dB
±2.0 dB
Reference
±2.0 dB
±2.0 dB
±2.0 dB
±2.0 dB
±2.0 dB
Reference
±2.0 dB
±2.0 dB
±2.0 dB
±2.0 dB
±2.0 dB
Reference
±2.0 dB
±2.0 dB
±2.0 dB
±2.0 dB
±2.0 dB
Reference
±2.0 dB
±2.0 dB
±2.0 dB
±2.0 dB
±2.0 dB
Reference
±2.0 dB
±2.0 dB
±2.0 dB
3
Performance Test
3-65