Harmonics, Spurious, and Sub-Harmonic Content – Microwave Output (96270A)
Use a high performance spectrum analyzer to make the measurements shown in
Tables A-24 and A-25.
Table A-24. 96270A Microwave Output Harmonics Test
Output
Level
Frequency
20 kHz
2.74 MHz
5.5 MHz
11 MHz
22 MHz
50 MHz
88 MHz
125 MHz
250 MHz
354 MHz
500 MHz
+16 dBm
1 GHz
1.25 GHz
1.4 GHz
1.8 GHz
2 GHz
2.7 GHz
5 GHz
9 GHz
10 GHz
12 GHz
13 GHz
nd
2
Harmonic
Freq
Limit
40 kHz
-60 dBc
5.48 MHz
-60 dBc
11 MHz
-60 dBc
22 MHz
-60 dBc
44 MHz
-60 dBc
100 MHz
-60 dBc
176 MHz
-60 dBc
250 MHz
-60 dBc
500 MHz
-60 dBc
708 MHz
-60 dBc
1 GHz
-60 dBc
2 GHz
-60 dBc
2.5 GHz
-60 dBc
2.8 GHz
-55 dBc
3.6 GHz
-55 dBc
4 GHz
-55 dBc
5.4 GHz
-55 dBc
10 GHz
-55 dBc
18 GHz
-55 dBc
20 GHz
-55 dBc
24 GHz
-55 dBc
26 GHz
-55 dBc
rd
3
Actual
Freq
60 kHz
8.22 MHz
16.5 MHz
33 MHz
66 MHz
150 MHz
264 MHz
375 MHz
750 MHz
1.062 GHz
1.5 GHz
3 GHz
3.75 GHz
4.2 GHz
5.4 GHz
8 GHz
8.1 GHz
15 GHz
-
-
-
-
Performance Test Record
Performance Test Record
Harmonic
Limit
Actual
-60 dBc
-60 dBc
-60 dBc
-60 dBc
-60 dBc
-60 dBc
-60 dBc
-60 dBc
-60 dBc
-60 dBc
-60 dBc
-60 dBc
-60 dBc
-55 dBc
-55 dBc
-55 dBc
-55 dBc
-55 dBc
-
-
-
-
-
-
-
-
A
A-25