Fluke 96000 Series Service Manual page 175

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

Modulation (96040A and 96270A, Optional)
Use a high performance measuring receiver to make the measurements shown
in Tables A-6, A-7, A-8, and A-9.
Level
+10 dBm
30 MHz
Level
125 MHz
+14 dBm
1 GHz
Level
62.5 MHz
+13 dBm
62.5 MHz
Level
125 MHz
+13 dBm
1 GHz
Table A-6. Amplitude Modulation Rate Test
Modulation
Frequency
Rate
20 kHz
100 kHz
Table A-7. Amplitude Modulation Depth Test
Modulation
Frequency
Rate
1 kHz
100 kHz
1 kHz
20 kHz
Table A-8. Frequency Modulation Rate Test
Modulation
Frequency
Rate
400 Hz
300 kHz
Table A-9. Frequency Modulation Deviation Test
Modulation
Frequency
Rate
400 Hz
100 kHz
400 Hz
100 kHz
400 Hz
300 kHz
Depth
Actual
50 %
50 %
Depth
Actual
90 %
90 %
90 %
90 %
Deviation
Actual
30 kHz
1 MHz
Deviation
Actual
10 kHz
10 kHz
300 kHz
300 kHz
1 MHz
1 MHz
Performance Test Record
Performance Test Record
Tolerance
±1 Hz
±10.0 Hz
Tolerance
±2.5 %
±2.5 %
±2.5 %
±2.5 %
Tolerance
±0.11 Hz
±10 Hz
Tolerance
±300 Hz
±300 Hz
±9.0 kHz
±9.0 kHz
±30 kHz
±30 kHz
A
A-5

Advertisement

Table of Contents
loading

This manual is also suitable for:

96040a96270a