Fluke 96000 Series Service Manual page 178

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

96000 Series
Users Manual
-7 dBm
-17 dBm
-27 dBm
A-8
Table 12. Level Accuracy (50 Ω) Test (cont).
Level
Frequency
1 kHz
20 kHz
100 kHz
10 MHz
100 MHz
300 MHz
1.4 GHz
3 GHz
4 GHz
1 kHz
20 kHz
100 kHz
10 MHz
100 MHz
300 MHz
1.4 GHz
3 GHz
4 GHz
1 kHz
20 kHz
100 kHz
10 MHz
100 MHz
300 MHz
1.4 GHz
3 GHz
4 GHz
Actual
Tolerance
±0.03 dB
±0.03 dB
±0.03 dB
±0.05 dB
±0.05 dB
±0.07 dB
±0.2 dB
±0.3 dB
±0.3 dB
±0.03 dB
±0.03 dB
±0.03 dB
±0.05 dB
±0.05 dB
±0.07 dB
±0.2 dB
±0.3 dB
±0.3 dB
±0.03 dB
±0.03 dB
±0.03 dB
±0.05 dB
±0.05 dB
±0.07 dB
±0.2 dB
±0.5 dB
±0.5 dB

Advertisement

Table of Contents
loading

This manual is also suitable for:

96040a96270a

Table of Contents