Level Accuracy - Microwave Output (96270A)
Use an AC Measurement Standard, a power meter and power sensors, and a
high performance spectrum analyzer/measuring receiver to make the
measurements shown in Tables A-20 and A-21.
Table A-20. 96270A Microwave Output Level Accuracy Test
Level
+16 dBm
+13 dBm
+3 dBm
Frequency
1 kHz
100 kHz
10 MHz
100 MHz
1 GHz
4.024 GHz
10 GHz
15 GHz
20 GHz
26.4 GHz
1 kHz
100 kHz
10 MHz
100 MHz
1 GHz
4.024 GHz
10 GHz
15 GHz
20 GHz
26.4 GHz
1 kHz
100 kHz
10 MHz
100 MHz
1 GHz
4.024 GHz
10 GHz
15 GHz
20 GHz
26.4 GHz
Performance Test Record
Performance Test Record
Actual
Typical Tolerance
±0.5 dB
±0.5 dB
±0.5 dB
±0.5 dB
±0.5 dB
±1.0 dB
±1.0 dB
±1.0 dB
±1.0 dB
±1.0 dB
±0.5 dB
±0.5 dB
±0.5 dB
±0.5 dB
±0.5 dB
±1.0 dB
±1.0 dB
±1.0 dB
±1.0 dB
±1.0 dB
±0.5 dB
±0.5 dB
±0.5 dB
±0.5 dB
±0.5 dB
±1.0 dB
±1.0 dB
±1.0 dB
±1.0 dB
±1.0 dB