The insertion loss of the 75
must be taken into account at each adjustment frequency.
11. Set the Power Meter for 0.001 dB resolution.
The next steps make the high and mid-level flatness adjustment of
the 75
12. Set the UUT to OPER.
13. Enter the measurement frequency corresponding to each Target into the
The recommended Power Meter has Power Sensor calibration
factors stored internally in the sensor. Setting the input frequency in
the Power Meter automatically selects the appropriate cal factor for
the measurement. However, the Power Sensor may also be
accompanied by a table of calibration factors which must be applied
externally.
The Power Meter autorange boundary may coincide with a
measurement point. To avoid unwanted Power Meter autoranging
during the adjustment process, manually lock the Power Meter range
when the UUT indicates the current adjustment point is a Transfer
Point and maintain the Power Meter range lock for all frequencies at
that amplitude. When the sequence reaches the next Transfer Point
allow the Power Meter to autorange and then manually lock the
Power Meter range again, continuing in this fashion for all points
requiring use of the Power Meter.
14. Using the blue arrow keys and rotary knob, edit the UUT Actual output until
15. Push Next Target.
16. Repeat steps 13 through 15 for all points requiring the use of the Power
If using a Power Sensor with different characteristics than the one
specified, it may be necessary to change sensors prior to this point.
Consult the Power Sensor manufacturer's manual for input signal
level specifications.
17. Disconnect the Power Meter and sensor from the UUT 75 Ω Leveling Head.
The next steps make the low-level flatness adjustment of the 75
Leveling Head using a Spectrum Analyzer/Measuring Receiver.
4-12
75 Ω Power Sensor using a Type-N female-to-female adapter. See Figure 3-
11 for equipment connections.
Ω
Leveling Head using a Power Meter and sensor.
Power Meter to enable cal factor (frequency) correction.
the Power Meter indicates the Target value stated. Push Accept Target.
Meter and sensor (the UUT text box will indicate Use Power Meter). Set the
UUT to STBY.
Connect the 75 Ω Leveling Head to the input of the Spectrum Analyzer via a
75 Ω to 50 Ω impedance-matching pad at the Spectrum Analyzer RF input.
Note
Ω
Type-N female-to-female adapter
Note
Note
Note
Note
Note
Ω