Service Manual
11. Repeat steps 5 through 10 for all output and harmonic frequencies in
12. Set the UUT as follows:
13. Set the Spectrum Analyzer as follows:
14. On the Spectrum Analyzer, push MARKER → PEAK.
15. Set the Spectrum Analyzer as follows:
16. Using the Spectrum Analyzer delta marker functions, measure the amplitude
Refer to the notes in steps (4) and (8) above regarding avoiding
overloading the spectrum analyzer mixer and observing and
measuring interharmonic signals on the analyzer display at or below
the tolerance limits with the above analyzer settings.
17. Repeat steps 13 through 16 for all output, sub harmonic, and interharmonic
3-72
Table 3-29, using the settings and limits shown for each.
Frequency
Level
Output
FREQ
FREQ
on the Spectrum Analyzer for the UUT's interharmonic relative to the
fundamental. The displayed (delta marker) value must be within the
tolerance listed in Table 3-30 for interharmonics.
frequencies in Table 3-30, using the settings and limits shown for each.
4.05 GHz
+16 dBm
OPER
4.05 GHz (fundamental)
6.075 GHz (interharmonic)