Fluke 96000 Series Service Manual page 187

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

Attenuation Accuracy - 75
Use a high performance measuring receiver to make the measurements shown
in Table A-18, relative to +10 dBm.
Frequency
Attenuation
0 dB
-3 dB
-13 dB
-23 dB
-33 dB
-43 dB
10 MHz
-53 dB
-63 dB
-73 dB
-82 dB
-91 dB
-101 dB
50 MHz
Ω
Levelling Head (96040A and 96270A)
Table A-18. Attenuation Accuracy (75 Ω) Test
Actual
---
0 dB
---
-3 dB
-13 dB
-23 dB
-33 dB
-43 dB
-53 dB
-63 dB
-73 dB
-82 dB
-91 dB
-101 dB
Specification
Uncertainty
Reference
Reference
±0.07 dB
±0.07 dB
±0.07 dB
±0.07 dB
±0.1 dB
±0.1 dB
±0.1 dB
±0.2 dB
±0.2 dB
±0.2 dB
±0.4 dB (Typ)
Reference
Reference
±0.07 dB
±0.07 dB
±0.07 dB
±0.07 dB
±0.1 dB
±0.1 dB
±0.1 dB
±0.2 dB
±0.2 dB
±0.2 dB
±0.4 dB (Typ)
Performance Test Record
Performance Test Record
Tolerance
Reference
Reference
A
A-17

Advertisement

Table of Contents
loading

This manual is also suitable for:

96040a96270a

Table of Contents