Fluke 96000 Series Service Manual page 50

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

96000 Series
Service Manual
10 MHz Output from
Frequency Standard
11. Set the Spectrum Analyzer as follows:
12. Set the UUT frequency to 2.7 GHz and the output to OPER.
13. On the Spectrum Analyzer push the MKR (marker) key and then select
14. On the Spectrum Analyzer, push the NEXT key and select CNT RESOL
15. Allow the Spectrum Analyzer sweep to complete. Do a marker peak search
16. Note the spectrum analyzer frequency count result and check that the
3-8
Instrument
Figure 3-3. Equipment Connections - Frequency Accuracy Test Part B
PRESET
EXT REF
COUPLING
REF LVL
FREQ
SPAN
BW
SIGNAL COUNT from the marker menu.
0.1 HZ from the counter resolution menu.
(marker to peak), then allow another sweep to complete to obtain a signal
frequency result.
frequency count value is within the tolerance listed in Table 3-2.
Front
Rear
External
Reference
Frequency
Input
On
DC
+30 dBm
5 GHz
50 kHz
100 Hz
Spectrum Analyzer
RF Input
50
Leveling Head
hpn390.eps