Fluke 96000 Series Service Manual page 126

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

96000 Series
Service Manual
The Power Meter autorange boundary may coincide with a
measurement point. To avoid unwanted Power Meter autoranging
during the adjustment process, manually lock the Power Meter range
when the UUT indicates the current adjustment point is a Transfer
Point and maintain the Power Meter range lock for all frequencies at
that amplitude. When the sequence reaches the next Transfer Point
allow the Power Meter to autorange and then manually lock the
Power Meter range again, continuing in this fashion for all points
requiring use of the Power Meter.
12. Using the blue arrow keys and rotary knob, edit the UUT Actual output until
13. Push Next Target.
14. Repeat steps 11 through 13 for all points requiring the use of the Power
15. Disconnect the Power Meter and sensor from the UUT base unit. Connect the
16. Set the UUT to OPER.
17. Set the Spectrum Analyzer reference level, span, resolution and/or video
Depending on the actual configuration of the Spectrum
Analyzer/Measuring Receiver used, the settings required to make
the highest accuracy level measurements may vary. Consult the
Spectrum Analyzer manufacturer's manual for further information.
For each frequency it is important to maintain consistent
analyzer/receiver settings as the target amplitude changes. For any
given frequency the Analyzer setting can be to suit, but that setting
must be maintained for all amplitudes at that frequency after the
transfer point. In particular, analyzer attenuator, reference level and
resolution bandwidth settings must not change. Typically, there are 3
or 4 amplitudes per frequency point during which the Analyzer
settings must be static.
18. Using the blue arrow keys and rotary knob, edit the UUT Actual output until
19. Push Next Target.
20. Repeat steps 17 through 19 for all necessary points requiring the use of the
21. Disconnect the Spectrum Analyzer/Measuring Receiver from the UUT Base
4-6
the Power Meter indicates the Target value stated. Push Accept Target.
Meter and sensor (the UUT text box states, Use Power Meter). Set the UUT
to STBY.
UUT SMA output to the input of the Spectrum Analyzer/Measuring Receiver.
bandwidth as necessary to make a high-quality level measurement at each
applied frequency and level.
the Spectrum Analyzer indicates the Target value stated. Push Accept
Target.
Spectrum Analyzer to make level adjustments. When complete the UUT
adjustment point title will indicate AM Depth.
Unit front panel Head RF Output connector and fit the UUT 50 Ω Leveling
Head. Connect the Leveling Head output to the Spectrum
Analyzer/Measuring Receiver RF input.
Note
Note
Note