8. Connect the thermal Power Sensor RF input directly to the UUT front panel
Microwave Output connector. See Figure 3-16 for equipment connections.
Figure 3-16. Microwave Output Level Accuracy Tests, High Frequency Points
9. Set the UUT as follows:
Mode
Signal
Frequency
Level
Output
10. Configure the Power Meter to indicate readings in dBm. Enter the
measurement frequency (0.1 MHz) into the Power Meter to enable cal factor
(frequency) correction.
Leveled Sine
Microwave Output
100 kHz
+13 dBm
OPER
Performance Tests
Power Meter
3
Performance Test
hpn369.eps
3-59