Table 3-12. 50 Ω Leveling Head, High Frequency Level Accuracy Test Points, Diode Power Sensor
Amplitude
100 kHz
10 MHz
100 MHz
-7 dBm
300 MHz
1.4 GHz
3 GHz
4 GHz
100 kHz
10 MHz
100 MHz
-17 dBm
300 MHz
1.4 GHz
3 GHz
4 GHz
100 kHz
10 MHz
100 MHz
-27 dBm
300 MHz
1.4 GHz
3 GHz
4 GHz
100 kHz
10 MHz
100 MHz
-37 dBm
300 MHz
1.4 GHz
3 GHz
4 GHz
100 kHz
10 MHz
100 MHz
-47 dBm
300 MHz
1.4 GHz
3 GHz
4 GHz
(96040A and 96720A)
Frequency
Reading
P
2
P
3
P
3
P
3
P
3
P
3
P
3
P
2
P
3
P
3
P
3
P
3
P
3
P
3
P
2
P
3
P
3
P
3
P
3
P
3
P
3
P
2
P
3
P
3
P
3
P
3
P
3
P
3
P
2
P
3
P
3
P
3
P
3
P
3
P
3
Performance Tests
P
Tolerance
out
Reference
±0.05 dB
±0.05 dB
±0.07 dB
±0.20 dB
±0.30 dB
±0.30 dB
Reference
±0.05 dB
±0.05 dB
±0.07 dB
±0.20 dB
±0.30 dB
±0.30 dB
Reference
±0.05 dB
±0.05 dB
±0.07 dB
±0.20 dB
±0.30 dB
±0.50 dB
Reference
±0.05 dB
±0.05 dB
±0.07 dB
±0.20 dB
±0.30 dB
±0.50 dB
Reference
±0.05 dB
±0.05 dB
±0.07 dB
±0.20 dB
±0.30 dB
±0.50 dB