Fluke 96000 Series Service Manual page 87

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

Table 3-19. 75 Ω Leveling Head, Low Level Accuracy Test Points, Part A (96040A and 96270A)
Part
100 kHz
10 MHz
100 MHz
A
300 MHz
1.4 GHz
2 GHz
Frequency
Amplitude
-33 dBm
-53 dBm
-72 dBm
-81 dBm (P
-33 dBm
-53 dBm
-72 dBm
-81 dBm (P
-33 dBm
-53 dBm
-72 dBm
-81 dBm (P
-33 dBm
-53 dBm
-72 dBm
-81 dBm (P
-33 dBm
-53 dBm
-72 dBm
-81 dBm (P
-33 dBm
-53 dBm
-72 dBm
-81 dBm (P
Reading
Reference
Reference
P
±0.15 dB
rel
P
±0.20 dB
rel
)
P
±0.70 dB
-81
rel
Reference
Reference
P
±0.15 dB
rel
P
±0.20 dB
rel
)
P
±0.70 dB
-81
rel
Reference
Reference
P
±0.15 dB
rel
P
±0.20 dB
rel
)
P
±0.70 dB
-81
rel
Reference
Reference
P
±0.20 dB
rel
P
±0.20 dB
rel
)
P
±0.70 dB
-81
rel
Reference
Reference
P
±0.15 dB
rel
P
±0.20 dB
rel
)
P
±1.00 dB
-81
rel
Reference
Reference
P
±0.50 dB
rel
P
±0.50 dB
rel
)
P
±1.00 dB
-81
rel
Performance Tests
Performance Test
P
Tolerance
out
3
3-45

Advertisement

Table of Contents
loading

This manual is also suitable for:

96040a96270a

Table of Contents