Fluke 96000 Series Service Manual page 111

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

Instrument
10 MHz Output from
Frequency Standard
Figure 3-19. Equipment Connections - Microwave Output Frequency Accuracy Test Part B
11. Set the spectrum analyzer as follows:
PRESET
EXT REF
COUPLING
REF LVL
FREQ
SPAN
BW
12. On the Spectrum Analyzer push the MKR (marker) key and then select
SIGNAL COUNT from the marker menu. Push the NEXT key and select CNT
RESOL 0.1HZ from the counter resolution menu.
13. Allow the Spectrum Analyzer sweep to complete. Do a marker peak search
(marker to peak), then complete another sweep to obtain a signal frequency
result.
14. Note the Spectrum Analyzer frequency count result and check that the
frequency count value is within the tolerance listed in Table 3-28
15. Set the UUT output frequency and the spectrum analyzer center frequency to
the next frequency listed in part B of Table 3-28.
16. Repeat steps 13 to 15.
17. Set the UUT output to STBY. Remove all connections.
Front
Spectrum Analyzer
Rear
External
Reference
Frequency
Input
On
DC
+30 dBm
5 GHz
50 kHz
100 Hz
3
Performance Tests
Performance Test
RF Input
hpn392.eps
3-69

Advertisement

Table of Contents
loading

This manual is also suitable for:

96040a96270a